_DIRECTIONS FOR USING THE_ HAMMOND Wall Cabinet
MANUFACTURED BY _THE WAITE & BARTLETT MAN’F’G CO._, 143 East 23d
Street, New York.
TO USE THE GALVANIC CURRENT.
To use the galvanic current place the ends of the conducting cords
in the two binding posts marked =13-13=. Place the socket of current
selector (=20=) on the pin marked O, and turn the lever of selector
on pin No. =1=, or on any number, according to the number of cells
desired; at the same time see that all the switch levers are on their
blank buttons, except the one at left of current selector, which should
be on its left-hand button marked Constant.
If it is desired to use cells from any other part of the series, place
the socket on the pin from which it is desired to begin, and having
previously moved the lever ahead of this pin move it along until the
desired number of cells intervenes between socket and lever. The number
of cells in circuit when the socket is not on O, but on a pin with a
number, is the difference between the number of the pin and number on
which the lever rests, for instance, if socket is on pin =12= and lever
on pin =22=, then =10= cells are being used, the starting point being
=12= cells from any part of the series may thus be brought into the
circuit. If socket be placed on pin =39= and lever on pin =40= then the
=40=th cell would be in use only.
_Always keep the socket on the lowest number and the lever on the
highest, as putting socket ahead of lever would reverse the current._
Should, through any accident, a cell become disconnected or run down,
or should it get out of order from any cause, it is only necessary to
short-circuit the cells and throw the milli-ampere-meter in circuit
by adjusting the switch =10=. By placing the socket on the pin marked
O and selecting cell after cell with the lever, the defective cell is
easily found by the milli-ampere-meter failing to deflect when the pin
corresponding to the defect is reached.
TO OBTAIN THE INTERRUPTED GALVANIC CURRENT.
To obtain the interrupted galvanic current, move the switch at left
of selector on to button marked interrupted, then turn the switch
just above the vibrator on to its button marked =11=, by adjusting
the screws at =14= and =15= the proper degree of vibration will be
obtained. It is necessary that the point of the screws at =15= should
touch the plate with every vibration as the current passes through at
this point and unless it touches, connections would not be made.
TO MEASURE THE RESISTANCE OF THE PATIENT.