In accurate work it is advisable to avoid the use of gas jets or Bunsen
flames in the neighbourhood of the electrometer, as the flame gases are
strongly ionized and take some time to lose their conductivity. If
radio-active substances are present in the room, it is necessary to
enclose the wires leading to the electrometer in fairly narrow tubes,
connected with earth. If this is not done, it will be found that the
needle does not move at a constant rate, but rapidly approaches a steady
deflection where the rate of loss of charge of the electrometer and
connections, due to the ionization of the air around them, is balanced
by the current to be measured. This precaution must always be taken when
observations are made on the very penetrating rays from active
substances. These rays readily pass through ordinary screens, and ionize
the air around the electrometer and connecting wires. For this reason it
is impossible to make accurate measurements of small currents in a room
which is used for the preparation of radio-active material. In course of
time the walls of the room become radio-active owing to the
dissemination of dust and the action of the radio-active
emanations[107].
=63. Electrometer key.= For work with electrometers of high sensibility,
a special key is necessary to make and break from a distance the
connection of the quadrants with earth in order to avoid electrostatic
disturbances at the moment the current is to be measured. The simple key
shown in Fig. 16 has been found very satisfactory for this purpose. A
small brass rod _BM_, to which a string is attached, can be moved
vertically up and down in a brass tube _A_, which is rigidly attached to
a bent metal support connected with earth. When the string is released,
this rod makes contact with the mercury _M_, which is placed in a small
metal vessel resting on a block of ebonite _P_. The electrometer and
testing vessel are connected with the mercury. When the string is
pulled, the rod _BM_ is removed from the mercury and the earth
connection of the electrometer system is broken. On release of the
string, the rod _BM_ falls and the electrometer is again earthed. By
means of this key, which may be operated at any distance from the
electrometer, the earth connection may be made and broken at definite
intervals without any appreciable disturbance of the needle.
[Illustration: Fig. 16.]