STANDARDISING INSTRUMENTS.--For standardising latex by dilution all that is
required is an instrument which will preserve a perpendicular position
while floating in latex, will be sufficiently sensitive to indicate fairly
small differences in density of latex, and has one mark on its aerial
portion accurately indicating a density corresponding to the required
standard. On scientific grounds it can be demonstrated that such an
instrument as employed in common practice would not be strictly
accurate.[6] It is not proposed, in this section of the book, to discuss
such considerations.
[6] De Vries, "Archief voor de Rubbercultuur."
Instruments of this nature are represented by the "Metrolac" (originating
from the Rubber Growers' Association) and other similar recorders. They
generally consist of a submersible bulb with a projecting stem which is
graduated. The "Metrolac" differs from others in that the bulb is of
torpedo form (thus reducing "skin friction"), and the graduations on the
stem indicate actual weight of dry rubber per gallon instead of the
ordinary specific gravity figures.
Theoretical considerations to the contrary, it is found in actual practice
in Malaya and Ceylon that, although such instruments are naturally delicate
and require careful manipulation, they are of considerable practical value
and satisfy a definite requirement. Until an instrument of greater accuracy
and equal simplicity can be discovered all estates should regard the
possession of a few "Metrolacs" as essential.
The nature of their construction and the average conditions under which
they are used (and abused) make it impossible to rely upon their accuracy
indefinitely or for any long period. It is always recommended, therefore,
that there should be at least two instruments available, one of which may
be in daily use, while the other is kept in safe custody and only employed,
say, once a week for purposes of checking the accuracy or degree of
inaccuracy of the other. This can be done with reasonable approximity by
placing both instruments in a tall vessel containing well-mixed and diluted
latex. Instruments showing a marked degree of inaccuracy should not be
preserved; but in cases of necessity "Metrolacs" from estates belonging to
company members of the Rubber Growers' Association may be sent to the
laboratories for repair and adjustment.[7]
[7] This applies to the gilt brass instruments. As the result of experiment
the Rubber Growers' Association are now introducing glass instruments.
These are necessarily more fragile, but while unbroken can be relied on to
give correct readings.