In a dark room, with the light from a candle reflected up through the
cylinder, move the color-bar along with a jerking motion until both
compartments have the same depth of color. The number upon the scale
corresponding to the portion of the color-bar which is now under the
cylinder gives the percentage of hemoglobin. While comparing the two
colors, place the instrument so that they will fall upon the right and
left halves of the retina, rather than upon the upper and lower
halves; and protect the eye from the light with a cylinder of paper or
pasteboard. After use, clean the metal cylinder with water, and wash
the capillary tube with water, alcohol, and ether, successively.
Results with this instrument are accurate to within about 5 per cent.
A recent modification of the von Fleischl apparatus by Miescher gives
an error which need not exceed 1 per cent. It is, however, better
adapted to laboratory use than to the needs of the practitioner.
(2) The Sahli hemoglobinometer (Fig. 66) is an improved form of the
well known Gowers instrument. It consists of a hermetically sealed
comparison tube containing a 1 per cent. solution of acid hematin, a
graduated test-tube of the same {145} diameter, and a pipet of 20
c.mm. capacity. The two tubes are held in a black frame with a white
ground-glass back.
[Illustration: FIG. 66.--Sahli's hemoglobinometer.]
Place a few drops of decinormal hydrochloric acid solution in the
graduated tube. Obtain a drop of blood and draw it into the pipet to
the 20 c.mm. mark. Wipe off the tip of the pipet, blow its contents
into the hydrochloric acid solution in the tube, and rinse well. In a
few minutes the hemoglobin is changed to acid hematin. Place the two
tubes in the compartments of the frame, and dilute the fluid with
water drop by drop, mixing after each addition, until it has exactly
the same color as the comparison tube. The graduation corresponding to
the surface of the fluid then indicates the {146} percentage of
hemoglobin. Decinormal hydrochloric acid solution may be prepared with
sufficient accuracy for this purpose by adding 15 c.c. of the
concentrated acid to 985 c.c. distilled water. A little chloroform
should be added as a preservative.
This method is very satisfactory in practice, and is accurate to
within 5 per cent. The comparison tube is said to keep its color
indefinitely, but, unfortunately, not all the instruments upon the
market are well standardized.
(3) Dare's hemoglobinometer (Fig. 67) differs from the others in using
undiluted blood. The blood is allowed to flow by capillarity into the
slit between two small plates of glass. It is then placed in the
instrument and compared with different portions of a circular disc of
colored glass. The reading must be made quickly, before clotting takes
place. This instrument is easy to use, and is one of the most
accurate.