A Treatise on Meteorological Instruments: Explanatory of Their Scientific Principles, Method of Construction, and Practical UtilityNegretti, Enrico Angelo Lodovico
Science
A Treatise on Meteorological Instruments: Explanatory of Their Scientific Principles, Method of Construction, and Practical Utility
Negretti, Enrico Angelo Lodovico
Meteorological instruments; Meteorology; Scientific apparatus and instruments
Given 23.214 Next greater 23.456
22.974 Next less 22.974
------ ------
.240 Difference .482
.240
And ---- = .5
.482
Temperature opposite next less 199.0
-----
Temperature required 199.5
=====
A similar method of interpolation in taking out numerical quantities is
applicable to almost all tables; and should be practised with all those
given in this work.
_Example._--Thus, in Liverpool, lat. 53 deg. 30' N., the barometer reading
29.876 inches, its attached thermometer 55 deg., and the correction of the
instrument being + .015 (including index error, capillarity and capacity),
what temperature should be assigned for the boiling point marked on the
thermometer?
Observed barometer 29.876
Correction + .015
------
29.891
Correction for temperature - .074
------
Reduced reading 29.817
Factor from Table V. 1.00077
-------
208719
208719
29817
-----------
Equivalent for lat. 45 deg. 29.83995909
===========
In Table VI., 29.84 gives temperature 211 deg.86.