=79. Casella's Mercurial Minimum Thermometer.=--The general form and
arrangement of this instrument is shown in fig. 60. A tube with large
bore, _a_, has at the end a _flat glass diaphragm_ formed by the abrupt
junction of a small chamber, _b c_, the inlet to which at _b_ is larger
than the bore of the indicating tube. The result of this is that on
setting the thermometer, as described below, the contracting force of the
mercury in cooling withdraws the fluid in the indicating stem only; whilst
on its expanding with heat, the long column does not move, the increased
bulk of mercury finding an easier passage into the small pear-shaped
chamber attached.
[Illustration: Fig. 60.]
We believe that a small speck of air must be confined in the chamber, _b
c_, to act as a spring to start the mercury from the chamber in the act of
setting the thermometer. Were this air not present, the mercury would so
adhere to the glass that no amount of shaking could induce it to flow from
the chamber.
_To set the Instrument_, place it in a horizontal position, with the back
plate, _d_, suspended on a nail, and the lower part supported on a hook,
_e_. The bulb end may now be gently raised or lowered, causing the mercury
to flow slowly until the bent part, _a_, _is full_ and the chamber, _b c_,
_quite empty_. At this point the flow of mercury in the long stem of the
tube is arrested, _and indicates the exact temperature_ of the bulb or air
at the time. On an increase of temperature the mercury will expand into
the small chamber, _b c_; and a return of cold will cause its recession
from this chamber only, until it reaches the diaphragm, _b_. Any further
diminution of heat withdraws the mercury down the bore to whatever degree
the cold may attain, where it remains until farther withdrawn by increased
cold, or till re-set for future observation.
MAXIMA AND MINIMA THERMOMETERS.
=80. Rutherford's= arrangement for obtaining a complete instrument for the
registration of heat and cold was simply mounting a maximum thermometer
and a minimum thermometer upon the same frame or slab. Thus constructed,
they are often called "day and night" thermometers, though somewhat
inappropriately; for in temperate climates the temperature of the night
sometimes exceeds that of the day, notwithstanding the reverse is the
general law of temperature. Fig. 61 will explain the arrangement of
Rutherford's day and night thermometer.
[Illustration: Fig. 61.]