The carbon dioxide may be absorbed in a tube containing soda lime
(F) or in a Geissler bulb (F') containing a concentrated solution
of potassium hydroxide (Note 2). The tube (F) is a glass-stoppered
side-arm U-tube in which the side toward the evolution flask and one
half of the other side are filled with small, dust-free lumps of soda
lime of good quality (Note 3). Since soda lime contains considerable
moisture, the other half of the right side of the tube is filled with
small lumps of dry, dust-free calcium chloride to retain the moisture
from the soda lime. Loose plugs of cotton are placed at the top of
each arm and between the soda lime and the calcium chloride.
The Geissler bulb (F'), if used, should be filled with potassium
hydroxide solution (1 part of solid potassium hydroxide dissolved in
two parts of water) until each small bulb is about two thirds full
(Note 4). A small tube containing calcium chloride is connected with
the Geissler bulb proper by a ground joint and should be wired to the
bulb for safety. This is designed to retain any moisture from the
hydroxide solution. A piece of clean, fine copper wire is so attached
to the bulb that it can be hung from the hook above a balance pan, or
other support.
The small bottle (G) with concentrated sulphuric acid (sp. gr. 1.84)
is so arranged that the tube (f) barely dips below the surface. This
will prevent the absorption of water vapor by (F) or (F') and serves
as an aid in regulating the flow of air through the apparatus. (H) is
an aspirator bottle of about four liters capacity, filled with water;
(k) is a safety tube and a means of refilling (H); (h) is a screw
clamp, and (K) a U-tube filled with soda lime.
[Note 1: The air current, which is subsequently drawn through the
apparatus, to sweep all of the carbon dioxide into the absorption
apparatus, is likely to carry with it some hydrochloric acid from
the evolution flask. This acid is retained by the silver sulphate
solution. The addition of concentrated sulphuric acid to this solution
reduces its vapor pressure so far that very little water is carried on
by the air current, and this slight amount is absorbed by the calcium
chloride in (E). As the calcium chloride frequently contains a small
amount of a basic material which would absorb carbon dioxide, it is
necessary to pass carbon dioxide through (E) for a short time and then
drive all the gas out with a dry air current for thirty minutes before
use.]
[Note 2: Soda-lime absorption tubes are to be preferred if a
satisfactory quality of soda lime is available and the number of
determinations to be made successively is small. The potash bulbs will
usually permit of a larger number of successive determinations without
refilling, but they require greater care in handling and in the
analytical procedure.]