Measurement of Unknown Pressure.--The potentiometer (fig. 610),
as ordinarily used, gives direct readings for voltages up to and
including 1.6 volts. For pressures higher than 1.6 volts, a volt
box or multiplier should be used. After obtaining the standard cell
balance, as previously described, place the double throw switch in
the position marked E.M.F. The balance for the unknown E.M.F. is
obtained by manipulating the tenths switch and rotating the contact
on the extended potentiometer wire. The final position of the two
contacts in conjunction with the position of the plug at the left of
the instrument indicates the voltage under test.
As directed above, use key R_{1} for rough adjustment, R_{2} for
intermediate adjustment, and key R_{0} for final adjustment.
Plug at 1 gives readings for voltage directly from settings of tenths
switch and extended wire contact.
Plug at .1 shunts the potentiometer circuit so that the voltage
measured is .1 of the reading taken directly from the scale. Hence,
the readings taken from the setting of the tenths switch and the slide
wire contact must be divided by 10.
To Balance Galvanometer for Unknown Voltage.--Place plug in hole 1
(fig. 610) for voltages up to 1.6, and in hole .1 for voltages up
to .16. Rotate the tenths switch until a condition of balance is
obtained exactly or approximately. To secure an exact balance, rotate
the contact on the extended wire. The unknown voltage can now be read
directly from the position of the tenths switch and the extended wire
contact if plug be at 1, or by dividing by 10 if plug be at .1.
EXAMPLE.--A balance was obtained with the tenths switch at 1.3, the
extended wire contact at 176 and the plug at 1. The voltage under
test, therefore, is 1.3176. If the plug at .1 had been used, the same
reading would have indicated .13176.
To ascertain if the current in the potentiometer circuit has altered
during a measurement, it is only necessary to plug in at 1, place
the double throw switch on STD. CELL and close the galvanometer key.
No deflection indicates that the current has not changed. If the
galvanometer deflect, the regulating rheostat must again be adjusted
until the galvanometer shows no deflection.
To Measure Voltages from 1.6 to 16.--Pressures up to 16 volts may
be measured by using a greater voltage across the BA posts (fig.
615). For this purpose a battery of about 20 volts should be used.
Insert the large plug at .1 and throw the switch to STD. CELL, then
balance the galvanometer by means of the regulating rheostat. When the
rheostat has been set to secure a balance, insert the large plug at
1, set the switch on E.M.F. and read the voltage in the usual manner.
Multiply the reading by 10.