(7) _Volumetric Solution._—It is advisable to substitute volumetric
solution for solution by filtration and washing, which is accomplished
by decomposing the substances in a graduated flask, the volume being
afterwards made up to the mark with distilled water after cooling. The
solution is then filtered without washing, and by means of a pipette an
aliquot part of the original volume is taken for precipitation. Thus
all retardations in the process are avoided, and likewise the chances
of error from washing on the filter. It is true that this method may
lead to a certain error due to the volume of the insoluble matter
which is left undecomposed, but since this insoluble matter is usually
small in quantity, and since it is always possible to diminish the
error therefrom by increasing the volume of the solution, this cause
of error is much less to be feared than those due to the difficulties
which may occur in the other method. Let us suppose, in order to
illustrate the above, that we are dealing with a phosphate containing
fifty per cent of insoluble sand which may be considered as an extreme
limit. In working on four grams of the material in a flask of 100
cubic centimeters capacity, there will be an insoluble residue of two
grams occupying a volume of about one cubic centimeter, the density of
the sand being generally nearly two. The one hundred cubic centimeter
flask will then contain only ninety-nine cubic centimeters of the real
solution, and the error at the most would be 0.01. This error could be
reduced to one-half by dissolving only two grams of the material in
place of four, or by making the volume up to 200 instead of 100 cubic
centimeters.
In general it may be said that the errors which do not exceed 0.01 of
the total matter under treatment, are negligible for all industrial
products. The method of volumetric solution does not present any
further inconvenience. It deserves to be and has been generally adopted
by reason of its rapidity in all the laboratories where many analyses
are to be made. In the volumetric method great care should be taken not
to make up to the volume until after the cooling to room temperature,
which may be speedily secured by immersing the flask in cold water.
Care should also be exercised in taking the sample for analysis by
means of the pipette immediately after filtration, and filtration
should take place as soon as the volume is made up to the standard. By
operating in this way the possible variations from changes of volume
due to changes of temperature are avoided.