=Daniell’s Pyrometer.=—In 1822 Daniell published an account
of a pyrometer based on the expansion of a platinum rod enclosed in
a plumbago tube. One end of the rod pressed against the end of the
tube, whilst the other end was free to move, and was connected to
a multiplying device which magnified the expansion, the increased
movement being indicated by a pointer, moving over a dial. The scale on
the dial was divided evenly into a suitable number of parts, it being
assumed that the difference between the expansions of graphite and
platinum was uniform at all temperatures. The scale was calibrated as
far as possible by comparison with a mercury thermometer, the remainder
being extrapolated. With this pyrometer Daniell obtained a value of
2233° F. for the melting point of silver, and 3479° F. for that of cast
iron—results considerably higher than those now accepted, but much
nearer than those obtained by Wedgwood. Daniell’s pyrometer was widely
used, and its modern representatives are fairly common. Platinum, owing
to its cost, is no longer used in these instruments, which are now
generally made with a graphite rod encased in an iron tube, on the end
of which the graduated dial is placed, as shown in fig. 69. Another
form, commonly used in baker’s ovens, is constructed with an iron rod
surrounded by a porcelain or fireclay tube.
[Illustration: FIG. 69.—EXPANSION PYROMETER.]
The defect of pyrometers of this type is that the coefficient of
expansion of the materials alters with prolonged heating, causing the
readings to become erroneous. Re-adjustment in boiling water or other
substance does not compensate properly for this alteration, as both
materials are not equally affected. Again, the readings will be too
low unless the whole of the expanding parts are in the interior
of the furnace, in which respect this pyrometer is inferior to a
thermo-electric instrument, which may be inserted at any convenient
depth, and may therefore be used for a greater variety of purposes.
The chief recommendation is cheapness; but an expansion pyrometer
should never be used for work of precision. A graphite rod in an iron
enclosure gives more consistent results than other materials.