=Management of Radiation Pyrometers.=—It is not advisable to
place a radiation pyrometer in the hands of an unskilled observer, as
intelligent oversight is required if good results are to be secured.
Care must be taken to adjust the galvanometer needle to zero before
taking a reading, and the needle should always be locked during
transit. When focusing on an object in a furnace it is necessary to
make certain that the red image seen is actually that of the object,
which may be done by moving the pyrometer until the side of the
object, or some special feature, is visible in the eye-piece, when
the pyrometer may be moved until the image surrounds the junction.
Occasions may arise, as in taking the temperatures of various zones
of a rotary cement-kiln or other furnace, in which it is required to
focus the mirror for a specified distance; in which case the author
has adopted the plan of placing a fixed pointer opposite the milled
head which controls the mirror (P, fig. 44) and focusing the bars of
a window at measured distances, marking the same on the milled head
opposite the pointer; and it would be a convenience if all radiation
pyrometers were thus marked initially. A good check to correct
focusing in the case of a heated object is to alter the focus in both
directions, and finally to adjust to the maximum reading, which should
correspond to the true focus.
Great care should be taken not to damage the mirror. If, in a workshop,
the surface become covered with dirt, this should be removed by gentle
brushing with a camel-hair brush or by blowing air over the mirror. The
focusing device should never be strained beyond its working limits;
when these are reached, the pyrometer should be moved bodily until the
object can be correctly sighted within the ordinary limits of the
movement of the milled head. If metallic fumes or dense smoke intervene
between the furnace and the pyrometer, the radiations will be impeded
and the temperature recorded will be too low; and in such cases the
pyrometer should be placed at the open end of a tube and sighted upon
the closed end, which should terminate at the spot under observation.