=Cambridge Optical Pyrometer.=—During the recent war the
manufacture of pyrometers of this type was taken up by the Cambridge
and Paul Instrument Company. The external form of the Cambridge optical
pyrometer is shown in fig. 58, in which an observer is shown using
the instrument, the accessories consisting of a 4-volt accumulator,
an ammeter, and an adjustable resistance for regulating the current
through the electric lamp used for comparison; and a standard
amyl-acetate lamp for adjusting the electric lamp to the correct
brightness. The scale is marked on a circular disc, and direct readings
are obtained from the position of a pointer which rotates with the
analyser. By interposing a monochromatic glass to dim the source, the
range of the pyrometer can be modified; and instruments are provided
in four ranges: 700°-1400° C.; 900°-2000° C.; 1200°-2500° C., and
1400°-4000° C.
The Cambridge optical pyrometer has proved a useful instrument in
skilled hands, and has been found of great service in the steel, glass,
and pottery industries. Trained observers have found it possible to
detect a difference of 10° C. at the region of 1900° C. The adjustment
of the two fields to equality, however, involves a judgment which
varies with different observers, and in practice it is advisable for
one individual to be entrusted to take all readings.
[Illustration: FIG. 58.—CAMBRIDGE OPTICAL PYROMETER.]
=Holborn-Kurlbaum Pyrometer.=—In the optical pyrometers
previously described a constant standard is used, and the brightness of
the light from the source varied until equality is obtained. The idea
of varying the brightness of the filament of an electric lamp until its
colour matched that of the source, and deducing the temperature from
the current taken by the lamp, was due to Morse, who used a filament
in the form of a flat spiral, heated by a battery of E.M.F. 40 volts.
This spiral was placed in a metal tube and interposed between the
eye and the heated object. The Holborn-Kurlbaum pyrometer, as made
by Siemens, is a refinement of that of Morse, and capable of reading
over a more extended range. In fig. 59, L is a small electric lamp
with a hairpin filament, as shown at A. This lamp is placed in a
telescope, so that the filament is in the focus of the eye-piece and is
lighted by a 4-volt accumulator, in series with which is a rheostat,
R, and a milliammeter, M. The heated source is focused by moving the
object-glass of the telescope, and both lamp and source are viewed
through red glass placed in front of the eye-piece, D. The rheostat,
R, is then adjusted until the tip of the filament is indistinguishable
from the background, which is illuminated by the source. If the lamp be
too bright, the filament will appear as a bright line; if duller than
the source, as a dark line; and when equal to the source it will merge
into the background. When equality is obtained, the milliammeter is
read, and the temperature deduced from the current taken by the lamp.