Suppose that it is required to measure with an electrometer the
ionization current between two horizontal metal plates _A_ and _B_ (Fig.
14) on the lower of which some active material has been spread. If the
saturation current is required, the insulated plate _A_ is connected
with one pole of a battery of sufficient E.M.F. to produce saturation,
the other pole being connected to earth. The insulated plate _B_ is
connected with one pair of quadrants of the electrometer, the other pair
being earthed. By means of a suitable key _K_, the plate _B_ and the
pair of quadrants connected with it may be either insulated or connected
with earth. When a measurement is to be taken, the earth connection is
broken. If the positive pole of the battery is connected with _A_, the
plate _B_ and the electrometer connections immediately begin to be
charged positively, and the potential, if allowed, will steadily rise
until it is very nearly equal to the potential of _A_. As soon as the
potential of the electrometer system begins to rise, the electrometer
needle commences to move at a uniform rate. Observations of the angular
movement of the needle are made either by the telescope and scale or by
the movement of the spot of light on a scale in the usual way. If the
needle is damped so as to give a uniform motion over the scale, the rate
of movement of the needle, _i.e._ the number of divisions of the scale
passed over per second, may be taken as a measure of the current through
the gas. The rate of movement is most simply obtained by observing with
a stop-watch the time taken for the spot of light, after the motion has
become steady, to pass over 100 divisions of the scale. As soon as the
observation is made, the plate _B_ is again connected with earth, and
the electrometer needle returns to its original position.
In most experiments on radio-activity, only comparative measurements of
saturation currents are required. If these measurements are to extend
over weeks or months, as is sometimes the case, it is necessary to adopt
some method of standardizing the electrometer from day to day, so as to
correct for variation in its sensibility. This is done most simply by
comparing the current to be measured with that due to a standard sample
of uranium oxide, which is placed in a definite position in a small
testing vessel, always kept in connection with the electrometer. Uranium
oxide is a very constant source of radiation, and the saturation current
due to it is the same from day to day. By this method of comparison
accurate observations may be made on the variation of activity of a
substance over long intervals of time, although the sensibility of the
electrometer may vary widely between successive measurements.