=63. Cog psychrometer.= This instrument, commonly called the
“egg-beater” psychrometer, has been devised to obviate certain
disadvantages of the sling psychrometer in field work, and has entirely
supplanted the latter in the writer’s own studies. It is smaller, more
compact, and the danger of breaking in carriage or in use is almost nil.
It has the great advantage of making it possible to take readings in a
layer of air less than two inches in thickness, and in any position.
Fairly accurate results can even be obtained from transpiring leaves.
The instrument can readily be made by a good mechanic, at a cost for
materials of $1.75, which is less than half the price for the sling
form. A single drawback exists in the use of short, Centigrade
thermometers, inasmuch as tables of relative humidity are usually
expressed in Fahrenheit. It is a simple matter, however, to convert
Centigrade degrees into Fahrenheit, mentally, or the difficulty may be
avoided by the conversion table shown on page 47, or by constructing a
Centigrade series of humidity tables. The fact that the wet and dry
bulbs revolve in the same path has raised a doubt concerning the
accuracy of the results obtained with this instrument. Repeated
comparisons with the sling psychrometer have not only removed this doubt
completely, but have also proved that the standardization of the
thermometers has been efficient.
=64. Construction and use.= A convenient form of egg-beater is the Lyon
(Albany, New York), in which the revolving plates can be readily
removed, leaving the axis and the frame. The thermometers used are of
the short Centigrade type. They are 4½ inches long and read from –5° to
50°. Eimer and Amend, 205 Third Ave., New York city, furnish them at 75
cents each. The thermometers are carefully standardized and compared,
and then grouped in pairs that read together. Each pair is used to
construct a particular psychrometer. Each thermometer is strongly wired
to one side of the frame, pieces of felt being used to protect the tube
and increase the contact. The frame is also bent at the base angles to
permit free circulation of air about the thermometer bulbs. The bulb of
one thermometer is covered with the proper cloth, and the psychrometer
is finished. Since the frame revolves with the thermometers, it is
necessary to pour the water on the wet bulb, or to employ a pipette or
brush. The thermometer bulbs are placed in the layer to be studied, and
the frame rotated at an even rate and with moderate rapidity. The
observation is further made as in the case of the sling psychrometer. As
the circle of rotation is less than three inches in diameter, and the
layer less than an inch, in place of nearly three feet for the sling
form, the instrument should not be moved at all for extremely localized
readings, but it must be moved considerably, a foot or more, if it is
desirable to obtain a more general reading.