This apparatus is designed much after the principle of the
Regnault-Reiset apparatus, in that there is a confined volume of air in
which the subject lives and which is purified by its passage through
vessels containing absorbents for water and carbon dioxide. Fresh oxygen
is added to this current of air and it is then returned to the chamber
to be respired. This principle, in order to be accurate for oxygen
determinations, necessitates an absolutely air-tight system and
consequently special precautions have been taken in the construction of
the chamber and accessories.
TESTING THE CHAMBER FOR TIGHTNESS.
As already suggested, the walls are constructed of the largest possible
sheets of copper with a minimum number of seams and opportunities for
leakage. In testing the apparatus for leaks, the greatest precaution is
taken. A small air-pressure is applied and the variations in height of a
delicate manometer noted. In cases of apparent leakage, all possible
sources of leak are gone over with soapsuds when there is a slight
pressure on the chamber. As a last resort, which has ultimately proven
to be the best method of testing, an assistant goes inside of the
chamber, it is then hermetically sealed, and a slight diminished
pressure is produced. Ether is then poured about the walls of the
chamber and the odor of ether soon becomes apparent inside of the
chamber if there is a leakage. Many leaks that could not be found by
soapsuds can be readily detected by this method.
VENTILATION OF THE CHAMBER.
The special features of the respiration chamber are the ventilating-pipe
system and openings for supplementary apparatus for absorption of water
and carbon dioxide. The air entering the chamber is absolutely dry and
is directed into the top of the chamber immediately above the head of
the subject. The moisture given off from the lungs and skin and the
expired gases all tend to mix readily with this dry air as it descends,
and the final mixture of gases is withdrawn through an opening near the
bottom of the chamber at the front. Under these conditions, therefore,
we believe we have a maximum intermingling of the gases. However, even
with this system of ventilation, we do not feel that there is
theoretically the best mixture of gases, and an electric fan is used
inside of the chamber. In experiments where there is considerable
regularity in the carbon-dioxide production and oxygen consumption, the
system very quickly attains a state of equilibrium, and while the
analysis of the outcoming air does not necessarily represent fairly the
actual composition of the air inside of the chamber, it evidently
represents to the same degree from hour to hour the state of equilibrium
that is usually maintained through the whole of a 6-hour experiment.
Public-domain text, read in full here on John Shaqi.