The operation of the instrument is as follows: As the cold blast used in
the apparatus would be useless for the working of the pyrometer if taken
direct from the cold blast main, owing to its irregularity of pressure,
the regulator that has been described is employed, and by its means an
absolutely steady flow of cold blast air at an unvarying pressure is
secured. The diameters of the inlet and outlet nozzles are so nicely
adjusted that, so long as both are at the same temperature, the outlet
nozzle, which is open to the atmosphere, will pass all the air that the
inlet nozzle can deliver without disturbing the pressure in the cistern,
O; but if heat be applied to the circulating air through the walls of the
spiral heater, the air expands in volume, and is unable to pass through
the outlet nozzle in its heated condition as rapidly as it is delivered
cold by the inlet nozzle. The consequence is that an increase of pressure
takes place in the apparatus between the two nozzles, and it is this
pressure that indicates the amount of heat that the air has taken up from
the hot blast pipe, in which the spiral heater is fixed. Then, again, as
this pressure is directly transmitted to the indicating liquid in the
cistern and the vertical tube immersed in it, a rise takes place in the
column which is in exact proportion to the expansion of the current of
blast passing through the spiral heater.
The method of graduating the indicator scales of the Frew pyrometer is
worthy of special notice. When the apparatus is fitted up, and before it
is permanently fixed in position, the spiral heater is placed in cold
water of known temperature, and the point noted at which the colored
liquid stands in the indicator tube. The water is then boiled, and the
rise in the liquid in the tube is again noted. Suppose, in the first
instance, the cold water temperature to be 62 deg. Fahr., and that, from
this point up to 212 deg. Fahr., the liquid to have risen 2¼ in. in the
tube; this is equal to 1½ in. per 100 deg. Fahr., and from these data a
scale is constructed, the correctness of which is easily verified by
transferring the spiral heater into an air bath or oil of high boiling
point, and then comparing the readings of the pyrometer scale with those
of a mercurial thermometer placed alongside of the spiral heater. By this
means it can be clearly demonstrated that, up to the highest point to
which it is safe to use a mercurial thermometer, the readings of the
pyrometer scale and that of the thermometer are identical.