As a determination of heat production is of considerable interest the
method will be given in some detail. Although the experiment sounds very
simple, it is actually somewhat difficult to carry out. The attainment
of temperature equilibrium between two solutions is very slow when one
wishes to obtain them to within 0.001° C. of the same temperature. After
many attempts, the following arrangement of apparatus (Fig. 33) was
found most satisfactory. About 10 c.c. luciferin solution was placed in
the inner tube (_D_) of a special non-silvered thermos bottle (_A_).
About 1 c.c. of luciferase solution was placed in a very thin-walled
glass tube (_E_) which was immersed in the luciferin solution and
connected with a small motor so that it could be slowly but constantly
rotated, thus stirring the solutions. Thermocouples (_L_ and _M_) of
advance (.008 in)--copper (No. 30, _B_ and _S_, enamel insulated) wire
were paraffined and placed in each tube and the copper wires connected
through a copper double throw switch (_C_) with a Leeds and Northrup
d'Arsonval wall galvanometer (No. 34637, silver strip suspension) of 35
ohms resistance and 310 megohms sensitivity. The constant temperature
junctions (_N_) were placed in a large Dewar flask (_B_) filled with
water at approximately the same temperature as the luciferin solution.
One mm. galvanometer scale division represented 0.003° C. and the
division readings could be estimated to tenths. By means of a glass rod
(_F_) placed in the tube containing luciferase solution, this tube
could be broken and the luciferase and luciferin solution mixed.
[Illustration: FIG. 33.--Apparatus for determining heat production
during luminescence of luciferin. A, special thermos tube. B, Dewar
flask for constant temperature junctions. C, double throw switch. D,
tube containing luciferin solution. E, tube containing luciferase
solution. F, glass rod for breaking E. G, rubber stopper with groove, K,
for pulley cord. H, cork closing thermos tube. J, brass sleeve in H
allowing rotation of E. L, thermojunction in luciferase solution. M,
thermojunction in luciferin solution. N, constant temperature
junctions.]
[Illustration: FIG. 34.--Curve showing temperature change when two tubes
containing water at the same temperature are mixed. 0.1 galvanometer
scale division = 0.003° C. Dots represent readings of thermocouple in
tube D; crosses readings of thermocouple in tube E.]