[26] _Pleuroscopes_.--As mentioned above the anterior commissure
laryngoscope and the esophageal specula make very efficient
pleuroscopes; but three different forms of pleuroscopes have been
devised by the author for pleuroscopy. The retrograde esophagoscope
serves very well for work through small fistulae.
_Measuring Rule_ (Fig. 7).--It is customary to locate esophageal
lesions by denoting their distance from the incisor teeth. This is
readily done by measuring the distance from the proximal end of the
esophagoscope to the upper incisor teeth, or in their absence, to the
upper alveolar process, and subtracting this measurement from the
known length of the tube. Thus, if an esophagoscope 45 cm. long be
introduced and we find that the distance from the incisor teeth to the
ocular end of the esophagoscope as measured by the rule is 20 cm., we
subtract this 20 cm. from the total length of the esophagoscope (45
cm.) and then know that the distal end of the tube is 25 cm. from the
incisor teeth. Graduation marks on the tube have been used, but are
objectionable.
[FIG. 7.--Measuring rule for gauging in centimeters the depth of any
location by subtraction of the length of the uninserted portion of the
esophagoscope or bronchoscope. This is preferable to graduations
marked on the tubes, though the tubes can be marked with a scale if
desired.]
_Batteries_.--The simplest, best, and safest source of current is a
double dry battery arranged in three groups of two cells each,
connected in series (Fig. 8). Each set should have two binding posts
and a rheostat. The binding posts should have double holes for two
additional cords, to be kept in reserve for use in case a cord becomes
defective.* The commercial current reduced through a rheostat should
never be used, because there is always the possibility of "grounding"
the circuit through the patient; a highly dangerous accident when we
consider that the tube makes a long moist contact in tissues close to
the course of both the vagi and the heart. The endoscopist should
never depend upon a pocket battery as a source of illumination, for it
is almost certain to fail during the endoscopy. The wires connecting
the battery and endoscopic instrument are covered with rubber, so that
they may be cleansed and superficially sterilized with alcohol. They
may be totally immersed in alcohol for any length of time without
injury.
* When this is done care is necessary to avoid attempting to use
simultaneously the two cords from one pair of posts.
[FIG 8.--The author's endoscopic battery, heavily built for
reliability.
It contains 6 dry cells, series-connected in 3 groups of 2 cells each.
Each group has its own rheostat and pair of binding posts.]