The graduated circle _a_ is horizontal and is divided directly to 15′,
the verniers enabling the readings to be carried further either to
single minutes, which is all that is usually necessary, or to
half-minutes in the cases of very perfect crystals. The divided circle
is rotated by means of the ring _b_ situated below, and the reading of
the verniers is accomplished with the aid of the microscopes _c_. The
circle which carries the verniers is not fixed, except when this is done
deliberately by means of the clamping screw _d_, but rotates with the
telescope _e_ to which it is rigidly attached by means of an arm and a
column _f_. A fine adjustment is provided with the clamping arrangement,
so that the telescope can be adjusted delicately with respect to the
divided circle. Both telescope and collimator are rigidly fixed at about
120° from each other during the actual measurements. The collimator _g_
is carried on a column _h_ definitely fixed to one of the legs (the back
one in Fig. 48) of the main basal tripod of the instrument. The signal
slit of the collimator is carried at the focus of the objective about
the middle of the tube _g_, the outer half of the latter being an
illumination tube carrying a condensing lens to concentrate the rays of
light from the goniometer lamp on the slit. The latter is not of the
usual rectilinear character, but composed of two circular-arc jaws, so
that, while narrow in the middle part like an ordinary spectroscope
slit, it is much broader at the two ends in order to be much more
readily visible; the central part is narrow in order to enable fine
adjustment to the vertical cross-wire of the telescope to be readily and
accurately carried out. The shape of this signal-slit will be gathered
from the images of the slit shown in Fig. 61 (page 126) in Chapter X.
The telescope carries an additional lens _k_ at its inner, objective,
end, in order that when this lens is rotated into position the telescope
may be converted into a low power microscope for viewing the crystal and
thus enabling its adjustment to be readily carried out.
The crystal _l_ is mounted on a little cone of goniometer wax (a mixture
of pitch and beeswax) carried by the crystal holder. The latter fits in
the top of the adjusting movements, which consist of a pair of
rectangularly arranged centring motions, and a pair of cylindrical
adjusting movements; the milled-headed manipulating centring screws of
the former are indicated by the letters _m_ and _n_ in Fig. 48, and
those which move the adjusting segments are marked _o_ and _p_. The top
screw fixes the crystal holder.