Handbook of anæstheticsRoss, J. Stuart (John Stuart)
Science
Handbook of anæsthetics
Ross, J. Stuart (John Stuart)
Anesthesia; Anesthetics
If the object be the removal of a tooth, the mask may now be removed,
and the dentist may rely upon a period of anæsthesia somewhat longer
than that furnished by pure nitrous oxide. He ought to secure
approximately one minute in which to do his work.
Anæsthesia by this apparatus may however be prolonged for an
indefinite time if desired. In order to maintain the patient in the
condition described above, it will be necessary gradually to increase
the supply of oxygen. For this purpose Hewitt added to his mixing
chamber a supplementary oxygen supply giving 10 or 20 volumes of
oxygen. As a matter of fact, all the necessary supply can be got
through the original ten holes _if the tension in the oxygen bag
be increased_. The regulation of all this requires, of course,
considerable practice and experience.
The signs upon which we rely for warning that the supply of oxygen is
insufficient to keep the patient safe, are chiefly the colour of the
face, which must not pass from dull pink to blue, and the size of the
pupil.
[Illustration: FIG. 19.--Diagrams to illustrate
action of (A) Hewitt (B) Teter gas oxygen apparatus. Note
that in A the mouths of both bags are guarded by valves of
inspiration, while in B the oxygen bag only possesses it.]
Deficiencies of Hewitt’s Apparatus.
These are chiefly two:--
(1) There is no means of producing _positive pressure_. Any
attempt to distend the nitrous oxide bag beyond a certain point
simply leads to escape of the gas through the mixing chamber and
out of the expiratory valve even during inspiration. To this
defect especially must we attribute the fact that an anæsthesia
deep enough for abdominal section is difficult to secure with
the Hewitt instrument.
(2) There is no means of securing _re-breathing_. The whole
administration must of necessity be conducted “upon the valves.”
This latter fault is remedied by the modification introduced by
_Burns_, who took away the inspiratory valve from the mouth of
the N_{2}O bag, and fitted a cap over the expiratory valve which
could be rotated so as to throw the valve out of action. The
author first met this modification at a Base Hospital in France,
and found it a great improvement upon the original instrument.
It is, however, only an imperfect attempt to adopt Teter’s chief
principle.
The Teter and Allied Machines.
The rapid spread of nitrous oxide and oxygen anæsthesia in the
U.S.A. brought forward a number of machines of which Teter’s was the
forerunner; the other well-known machine of the group is the Clarke.
They differ in principle from the Hewitt apparatus in that they permit
re-breathing and the use of positive pressure. Diagrammatically, the
two are contrasted in Fig. 19.