[Illustration: Fig. 520.--Graduation of tangent galvanometer scale with
divisions representing tangent values. In the figure let a tangent OT
be drawn to the circle, and along this line let any number of equal
divisions be set off, beginning at O. From these points draw lines back
to the center. The circle will thus be divided into a number of spaces, of
which those near O are nearly equal, but which get smaller and smaller as
they recede from O. These unequal spaces correspond to equal increments
of the tangent. If the scale were divided thus, the readings would be
proportional to the tangents.]
Ques. Upon what does the sensibility of a tangent galvanometer depend?
Ans. It is directly proportional to the number of turns of the coil and
inversely proportional to the diameter of the coil.
Ques. How may the tangent galvanometer be used as an ammeter?
Ans. The strength of the current may be calculated in amperes by the
formula given below when the dimensions of the instrument are known.
The needle is supposed to be subject to only the earth's magnetism and to
move in a horizontal plane. The current is calculated as follows:
(1) amperes = ((H × _r_)/N) tan _x_
in which
H = constant from table below;
r = radius of coil;
N = number of turns of coil;
x = angle of deflection of needle.
The constant H, given in the following table represents the horizontal
force of the earth's magnetism for the place where the galvanometer is
used. Each value has been multiplied by (2[pi])/10 so that the formula (1)
for amperes is correct as given.
Table of Galvanometer Constants.--Values of H.
Boston | .699|
Chicago | .759|
Denver | .919|
Jacksonville | 1.094|
London | .745|
Minneapolis | .681|
New York | .744|
New Haven | .731|
Philadelphia | .783|
Portland, Me. | .674|
San Francisco | 1.021|
St. Louis | .871|
Washington | .810|