Use of Mirrors in Galvanometers.--In order that small currents may be
measured accurately, some means must be provided to easily read a small
deflection of the needle. Accordingly, it is desirable that the pointer
be very long so that a large number of scale divisions may correspond to
small deflections. In construction, since sensitive galvanometers must
be made with the moving parts of little weight, it would not do to use a
long needle, hence a ray of light is used instead, which is reflected on
a distant scale by a small mirror attached to the moving part.
In the Thompson mirror reflecting galvanometer, as shown in fig. 528,
a small vertical slit is cut in the lamp screen below the scale, and
the ray of light from the lamp, passing through the slit, strikes
the mirror which is about three feet distant, and which reflects the
beam back to the scale. It should be noted that the angle between
the original ray of light and the reflected ray is twice the angle
of the deflection of the mirror; the deflections of the ray of light
on the scale, however, are practically proportional to the strength
of currents through the instrument. The mirror arrangement as shown
in fig. 528, requires a darkened room for its operation, but such is
not necessary when a telescope is used as in fig. 529. Here the scale
readings are reflected in the mirror and their value observed by the
telescope without artificial light.
[Illustration: Fig. 528.--Telescope method of reading galvanometer
deflections by reflection of scale reading in mirror. Here two mirrors
are used, but in most cases the telescope is pointed directly toward the
mirror on galvanometer shown in fig. 527, because the two mirror system,
as illustrated in the figure, is used on portable galvanometers since it
is the more compact.]
Damping.--This relates to the checking or reduction of oscillations.
Thus, a galvanometer is said to be damped when so constructed that any
oscillations of the pointer which may be started, rapidly die away.
Galvanometers are frequently provided with damping devices for the purpose
of annulling these oscillations, thus causing the moving part to assume
its final position as quickly as possible.
Sometimes the instrument is fitted with a damping coil, or closed coil so
arranged with respect to the moving system that the oscillations of the
latter give rise to electric currents in the closed coil, whereby energy
is dissipated. Again, air vanes are employed, but anything in the nature
of solid friction cannot be used.