=The Condenser= is so adjusted as to concentrate the rays of light about 2
mm above its surface, which is the average thickness of the microscopic
slides, under an angle equal to the angular aperture of the highest
power objective, which is about 120°.
In the lower power objectives the angular aperture is to be reduced by
the iris diaphragm.
To focus the Condenser properly it is adjustable by rack and pinion
movement on Stands A, I, Ia, Ib, and by side screw on Stands II and IIa.
4. =The body= of the microscope carries the tube and is fitted with two
adjustments for focussing; a coarse adjustment and a fine adjustment.
The smooth working of these adjustments is one of the chief requirements
of a good microscope.
=The coarse adjustment= consists of rack and pinion and moves the tube of
the microscope in a vertical groove with the least possible friction.
The teeth of the rack and pinion are set obliquely to the axis of the
tube and are so adjusted that two teeth of each are always interlocked,
thus avoiding all possibility of slipping or loss of motion in the
movement.
[Illustration: Figure of the Micrometer-screw of the fine adjustment.]
=The fine adjustment= consists of a micrometer screw which moves the tube
and arm of the microscope supporting the coarse adjustment vertically
upon a triangular column. This column is virtually a continuation upward
of the upright support of the base of the microscope. Around the column
is a broad collar so accurately fitted that it moves smoothly upon it
with a minimum of friction and still without lateral motion, the
movement being controlled by means of a micrometer screw at the top of
the column. The head of this micrometer screw, a section of which is
shown in the accompanying figure, is milled and is graduated in such a
way as to indicate the exact degree of motion of the microscope tube
which is accomplished by turning the micrometer screw, each division of
the graduation corresponding to a motion of the tube through 1/100
millimeter, and a complete revolution of the screw corresponding to a
motion of the tube through ½ millimeter. The accuracy of this adjustment
and the smoothness of its motion must be conceded to be a triumph of
mechanical skill.
=Stands A.= and =I.= are fitted with a =new fine adjustment= device (one
division on the micrometer screw corresponding to 1/1000 mm.) which is
fully described on pages 25-26.
5. =The tube= of the microscope encloses a second tube, or draw-tube,
which latter carries the eye-piece, and is so graduated as to indicate
the exact length of tube in use at any time. The lower end of the tube
has a thread into which the various objectives or a nose-piece screw.
Tube and objectives are provided with the so-called "Society screw".