By this adjustment of the lenses of the condenser it is possible to do
away with disturbing chromatic aberration. The rays of light leave the
condenser in moderate convergence.
For direct projection from the preparation the apparatus consists of the
optical bench upon which are mounted the various accessories in proper
succession (Fig. p. 88):
1. A large cooling cell, which is to be filled with water or alum
solution.
2. The bellows with tube. This is connected with the cooling cell by
means of a metal ring which slips over the end of the cell and is held
in place by a screw. The metal ring is attached to one end of the
bellows, while the other end carries a large tube.
3. The microscope condensers and diaphragm holder. A larger tube fits
over the tube of the bellows and carries on a revolving arm the cylinder
diaphragm and the two condensers. For low power objectives Nos. 1-3, the
cylinder diaphragm and the condenser of longer focus have to be used.
The latter at the same time serves as the objective for lantern slide
projection.
The other condenser is especially needed when high powers are employed
(Objectives 4 to 6). A small rack and pinion arrangement serves for the
adjustment of condenser and diaphragm.
4. The object stage and small cooler. The stage has connected with it a
second cooling cell on which the preparation rests. This arrangement
permits prolonged use of the condenser and the greatest intensity of
light without burning the most delicate specimens. The adjustment of the
object is accomplished by means of an arrangement which permits any
particular point to be brought accurately into the field.
[Illustration]
5. The objective carrier is provided with a nose-piece for three
objectives, with a broad projection tube, having an Iris diaphragm for
the purpose of narrowing the aperture, and with a narrow tube which
screws into the broad one, in case it is desired to use oculars in
making the projection. The adjustments of the objectives are
accomplished by rack and pinion and by micrometer screw.
In order to change readily from microscopic to lantern-slide projection,
the parts 4 and 5 have been provided with hinges, by means of which they
can be thrown to one side after loosening a lever (see illustration
below).
6. Wooden cover with cloth curtain which fits over the diaphragm
carrier, the stage and the objectives, to shut off any light which may
escape at the sides.
For the projection of large preparations of from 8-35 mm diameter our
photographic objectives of 24, 35, 42 and 64 mm focal distance are best
adapted (p. 86). In connection with these low powers the large tube only
should be used.
[Illustration]
With this projection apparatus the powers of microscope objectives till
number 6 may be well employed.