The dressings arranged in the large tray are placed in the sterilizer
and the supply of steam is maintained through the constant and steady
flow of water from the reservoir, which compensates the evaporation in
the sterilizer. In about twenty minutes the formation of steam in the
top of the reservoir exerts sufficient pressure to force all the boiling
water from the reservoir into the sterilizer to the depth of about 1½
inches. The tray of instruments is now inserted and the process continued
for another ten minutes. Much less heat is required with this apparatus
than with those of ordinary type, while sterilization can be continued
uninterruptedly for one and one half hours, if need be.
=Sprague Sterilizer.=—The most perfect sterilizer is that of the Sprague
type, in which a dry chamber is surrounded by steam under pressure. The
apparatus is shown in Fig. 15.
[Illustration: FIG. 15.—SPRAGUE TYPE OF STERILIZER.]
Its cylindrical chamber is surrounded by two heavy copper shells, the
space between which is occupied by the water. This compartment is
entirely shut off from the sterilizing chamber, and as the steam is
generated, the inner, or sterilizing, chamber becomes heated to a degree
nearly equal to that of the steam in the surrounding cylinder; this
prevents any condensation of steam taking place in the dressings. By
opening the lever-handled valve at the bottom of the sterilizer in the
rear, and the valve to the right, on top of the sterilizer, and allowing
them to remain open for a space of four or five minutes, a vacuum is
formed in the sterilizing chamber. These two valves are then closed, the
lower one first, and the steam from the outer cylinder is allowed to
enter the chamber, by opening the left valve on top.
The contents should be allowed to sterilize for twenty or twenty-five
minutes under a pressure of fifteen pounds. Then close the steam-supply
valve; open the vacuum valve (right) and the lever-handled valve at the
bottom; leave these open about the same time as in creating a vacuum at
the beginning of the process; close both valves, then open the air-filter
valve on the door, in order to break the vacuum; the door can then be
opened and the dressings be taken out dry and absolutely sterile.