considerably, the change in some cases representing
over 100° C. on the indicator. A stable condition, due to the relief of
strains or other change, was finally reached, and the conclusion drawn
that the materials should be thoroughly annealed before calibration. It
is desirable in all cases periodically to test the junctions at some
standard temperature, and if any conspicuous error be noted, to replace
the old junction by a new one.
In addition to the errors due to slow physical changes, a junction
may be altered considerably, if imperfectly protected, owing to the
chemical action of furnace gases, or of solids with which the junction
may come into contact. The vapours of metals such as lead or antimony
are very injurious; and platinum in particular is seriously affected
by vapours containing phosphorus, if in a reducing atmosphere. So
searching is the corrosive action of furnace gases that adequate
protection of the junction is essential if errors and damage are to be
avoided. When a wire has once been corroded, a junction made with it
will not develop the same E.M.F. as before.
=Electromotive Force developed by Typical Junctions.=—The
following table exhibits the E.M.F. generated by several junctions
for a range of 100° C., taken at the middle part of the working range
in each case. These figures are subject to considerable variation,
according to the origin of the metals.
──────────────────────────────────────────┬─────────────────────────
│ E.M.F. in millivolts for
Couple. │ a rise of 100° at middle
│ of working range.
──────────────────────────────────────────┼─────────────────────────
Platinum-rhodioplatinum (10 per cent. Rh) │ 1·1
Platinum-iridioplatinum (10 per cent. Ir) │ 1·2
Nickel-constantan │ 2·3
Copper-constantan │ 5·8
Nickel-copper │ 6·1
Iron-constantan │ 6·7
Hoskin’s alloys │ 7·4
──────────────────────────────────────────┴─────────────────────────
It will be noted that the base-metal junctions give much higher values
than the platinum series, and hence can be used with a less sensitive,
and therefore cheaper, indicator. Base-metal junctions are also, in
consequence of the greater E.M.F. furnished, capable of yielding more
sensitive readings over a selected range of temperature.
[Illustration: FIG. 4.—PRACTICAL FORM OF THERMO-ELECTRIC PYROMETER.]