The following simplified method of procedure is suggested for
subjective testing with the Ski-optometer, although as previously
explained, the refractionist may employ his customary method,
overcoming the annoyance of transferring trial-case lenses and the
setting of each cylinder individually. The Ski-optometer has been
constructed and based upon the golden rule of refraction: “As much plus
or as little minus spherical, combined with as little minus cylinder
power as the patient accepts.”
By applying this rule as in the above method and starting with +5.D.
spherical, watching the two zeros (Fig. 4) and rapidly reducing +1.25D
each time, we will assume that +1.25D gives 20/30 vision; as a final
result +1.D. will possibly give 20/25 vision.
The patient’s attention should next be directed to the most visible
line of type, preferably concentrating on the letter “E” or the clock
dial chart—either of which will assist in determining any possible
astigmatism. Since the Ski-optometer contains concave cylinders
exclusively, the next move should be the setting of its axis indicator
at 180°, commonly understood as “with the rule.” One should then
proceed to determine the cylinder lens strength by turning the reel
containing the cylindrical lenses (Fig. 8). Should the patient’s vision
fail to improve after the -.50D. cylinder axis 180° has been employed,
the refractionist, in seeking an improvement, should then slowly move
the axis indicator through its entire arc.
With the cylinder added, regardless of axis, poor vision might indicate
the absence of astigmatism. If astigmatism exists, vision will usually
show signs of improvement at some point, indicating the approximate
axis. Once the latter is ascertained, the refractionist may readily
turn the Ski-optometer’s cylinder reel and obtain the correct cylinder
lens strength, after which the axis indicator should be moved in either
direction in order to obtain the best possible vision for the patient.
The refractionist should always aim to obtain normal (or 20/20) vision
with the weakest concave cylinder, combined with the strongest plus
sphere, or weakest minus sphere.
PROCEDURE FOR USING MINUS CYLINDERS EXCLUSIVELY
For the benefit of those who have never used minus cylinders
exclusively in making their examinations, we will assume that the
patient requires O.U. +1.D sph. = -1D cyl. axis 180° for final
correction; the latter, in its transposed form, being equivalent to
+1.D. cylinder axis 90°. Unquestionably the best method is the one that
requires the least number of lens changes to secure the final result.
To obtain this, the following order of lens change should be made:
First, +1.D. sphere is finally determined and allowed to remain in
place. Concave cylinders are then employed in quarters until the final
results of +1.D. spherical, combined with -1.D. cylinder axis 180° is
secured. This necessitates the change of but _four_ cylindrical lenses
as shown in routine “A” as follows: