For use in measuring the temperature of the air and of the copper wall
of the calorimeters, as well as the rectal temperature of the subject, a
series of resistance thermometers is employed. These are so connected
on the observer's table that they may be brought into connection with
two Wheatstone bridges, W_{1} and W_{2}. Bridge W_{1} is used for the
resistance thermometers indicating the temperature of the wall and the
air. Bridge W_{2} is for the rectal thermometer. Since similar
thermometers are inserted in both calorimeters, it is necessary to
introduce some switch to connect either set at will and hence the
double-throw switches S_{1}, S_{2}, and S_{3} allow the use of either
the wall, air, or rectal thermometer on either the bed or chair
calorimeter at will. Since the bridge W_{1} is used for measuring the
temperature of both the wall and the air, a fourth double-pole switch,
S_{4}, is used to connect the air and wall thermometers alternately. The
double-contact key, K_{1}, is connected with the bridge W_{1} and is so
arranged that the battery circuit is first made and subsequently the
galvanometer circuit. A similar arrangement in K_{2} controls the
connections for the bridge W_{2}.
GALVANOMETER.
The galvanometer is of the Deprez-d'Arsonval type and is extremely
sensitive. The sensitiveness is so great that it is desirable to
introduce a resistance of some 500 ohms into the thermal-junction
circuits. This is indicated at the top of the diagram near the
galvanometer. The maximum sensitiveness of the galvanometer is retained
when the connection is made with the Wheatstone bridges. The
galvanometer is suspended from the ceiling of the calorimeter laboratory
and is free from vibration.
RESISTANCE FOR HEATING COILS.
Public-domain text, read in full here on John Shaqi.