Standard methods for the examination of water and sewageAmerican Public Health Association. Laboratory Section
Science
Standard methods for the examination of water and sewage
American Public Health Association. Laboratory Section
Sewage -- Analysis; Water -- Analysis
REPRESENTATIVE SAMPLES.
Care should be taken to obtain a sample that is truly representative of
the liquid to be analyzed. With sewages this is especially important
because marked variations in composition occur from hour to hour.
Satisfactory samples of some liquids can be obtained only by mixing
together several portions collected at different times or at different
places—the details as to collection and mixing depending upon local
conditions.
PHYSICAL EXAMINATION.
TEMPERATURE.
The temperature of the sample, if taken, shall be taken at the time of
collection, and shall be expressed preferably in degrees Centigrade, to
the nearest degree, or closer if more precise data are required. The
thermophone[109] is recommended for obtaining the temperature of water
at various depths below the surface.
TURBIDITY.
The turbidity of water is due to suspended matter, such as clay, silt,
finely divided organic matter, microscopic organisms, and similar
material.
TURBIDITY STANDARD.[110]
The standard of turbidity shall be that adopted by the United States
Geological Survey, namely, a water which contains 100 parts per million
of silica in such a state of fineness that a bright platinum wire 1
millimeter in diameter can just be seen when the center of the wire is
100 millimeters below the surface of the water and the eye of the
observer is 1.2 meters above the wire, the observation being made in the
middle of the day, in the open air, but not in sunlight, and in a vessel
so large that the sides do not shut out the light so as to influence the
results. The turbidity of such water is arbitrarily fixed at 100 parts
per million.
For preparation of the silica standard dry Pear’s “precipitated fuller’s
earth” and sift it through a 200–mesh sieve. One gram of this
preparation in 1 liter of distilled water makes a stock suspension which
contains 1,000 parts per million of silica and which should have a
turbidity of 1,000. Test this suspension, after diluting a portion of it
with nine times its volume of distilled water, by the platinum wire
method to ascertain if the silica has the necessary degree of fineness
and if the suspension has the necessary degree of turbidity. If not,
correct by adding more silica or more water as the case demands.[A]
Footnote A:
This method of correction very slightly alters the coefficient of
fineness of the standard, but does not noticeably affect its use.
Standards for comparison shall be prepared from this stock suspension by
dilution with distilled water. For turbidity readings below 20,
standards of 0, 5, 10, 15, and 20 shall be kept in clear glass bottles
of the same size as that containing the sample; for readings above 20,
standards of 20, 30, 40, 50, 60, 70, 80, 90, and 100 shall be kept in
100 cc. Nessler tubes approximately 20 millimeters in diameter.