Sterilisation by hot air is effected in the hot-air oven (Fig. 18). This
is a rectangular, double-walled metal box, mounted on a stand and heated
from below by a large Bunsen burner. The interior of the oven is
provided with loose shelves upon which the articles to be sterilised are
arranged, either singly or packed in square wire baskets or crates, kept
specially for this purpose. One of the sides is hinged to form a door.
The central portion of the metal bottom, on which the Bunsen flame would
play, is cut away, and replaced by firebrick plates, which slide in
metal grooves and are easily replaced when broken or worn out. The top
of the oven is provided with a perforated ventilator slide and two
tubulures, the one for the reception of a centigrade thermometer
graduated to 200° or 250°C., the other for a thermo-regulator. An
ordinary mercurial thermo-regulator may be used but it is preferable to
employ a regulating capsule of the Hearson type (see p. 219) with a
spring arm adjusted to the lever so that when the boiling-point of the
capsule (e. g., 175°C.) is reached the gas supply is absolutely cut
off and the jet cannot again be lighted until the spring-arm has been
readjusted by hand. The thermo-regulator is by no means a necessity, and
may be replaced by a large bore thermometer with a sliding platinum
point, connected with an electric bell, which can be easily adjusted to
ring at any given temperature. Even if the steriliser is provided with
the capsule regulator above described the contact thermometer should
also be fitted.
[Illustration: FIG. 26.--Hot-air oven.]
TO USE THE HOT-AIR OVEN.--
1. Place the crates of test-tubes, metal cases containing plates and
pipettes, loose apparatus, etc., inside the oven, taking particular care
that none of the cotton-wool plugs are in contact with the walls,
otherwise the heat transmitted by the metal will char or even flame
them.
To prepare a wire crate for the reception of test-tubes,
etc., cover the bottom with a layer of thick asbestos cloth;
or take some asbestos fibre, moisten it with a little water
and knead it into a paste; plaster the paste over the bottom
of the crate, working it into the meshes and smoothing the
surface by means of a pestle. When several crates have been
thus treated, place them inside the hot-air oven, close the
door, open the ventilating slide, light the gas, and run the
temperature of the interior up to about 160° C. After an
interval of ten minutes extinguish the gas, open the oven
door, and allow the contents to cool. The asbestos now forms
a smooth, dry, spongy layer over the bottom, which will last
many months before needing renewal, and will considerably
diminish the loss of tubes from breakage.
Copper cylinders and large test-tubes intended for the
reception of pipettes are prepared in a similar manner, in
order to protect the points of these articles from injury.
Public-domain text, read in full here on John Shaqi.
The Elements of Bacteriological Technique: A Laboratory Guide for Medical, Dental, and Technical Students. Second Edition Rewritten and Enlarged. — John Shaqi
Reviews
Reviews
No reviews yet
Be the first to share your thoughts on this work.
Join the Discussion
Join the discussion
Sign in to leave a comment or review.
Sign InorCreate an account