The Elements of Bacteriological Technique: A Laboratory Guide for Medical, Dental, and Technical Students. Second Edition Rewritten and Enlarged.Eyre, J. W. H. (John William Henry)
Science
The Elements of Bacteriological Technique: A Laboratory Guide for Medical, Dental, and Technical Students. Second Edition Rewritten and Enlarged.
Eyre, J. W. H. (John William Henry)
Bacteriology -- Technique
_Superheated steam_--i. e., steam under pressure (see
Pressure-temperature table, Appendix, page 500) in sealed vessels at a
temperature of 115° C.--will destroy both the vegetative and the sporing
forms of bacteria within fifteen minutes; if the pressure is increased,
and the temperature raised to 120° C., the same end is attained in ten
minutes. This method was formerly employed for the sterilisation of
media (and indeed is so used in some laboratories still), but most
workers now realise that media subjected to this high temperature
undergo hydrolytic changes which render them unsuitable for the
cultivation of the more delicate micro-organisms. The use of superheated
steam should be restricted almost entirely to the disinfection of such
contaminated articles, old cultivations, etc., as cannot be dealt with
by dry heat or the actual furnace. Sterilisation by means of superheated
steam is carried out in a special boiler--Chamberland's autoclave (Fig.
30). The autoclave consists of a stout copper cylinder, provided with a
copper or gun-metal lid, which is secured in place by means of bolts and
thumbscrews, the joint between the cylinder and its lid being
hermetically sealed by the interposition of a rubber washer. The cover
is perforated for a branched tube carrying a vent cock, a manometer, and
a safety valve. The copper boiler is mounted in the upper half of a
cylindrical sheet-iron case--two concentric circular rows of Bunsen
burners, each circle having an independent gas-supply, occupying the
lower half. In the interior of the boiler is a large movable wire
basket, mounted on legs, for the reception of the articles to be
sterilised.
TO USE THE AUTOCLAVE.--
1. Pack the articles to be sterilised in the wire basket.
2. Run water into the boiler to the level of the bottom of the basket;
also fill the contained flasks and tubes with water.
3. See that the rubber washer is in position, then replace the cover and
fasten it tightly on to the autoclave by means of the thumbscrews.
4. Open the vent cock and light both rings of burners.
5. When steam is issuing in a steady, continuous stream from the vent
tube, shut off the vent cock and extinguish the outer ring of gas
burners.
6. Wait until the index of the manometer records a temperature of 120°
C., then regulate the gas and the spring safety valve in such a manner
that this temperature is just maintained, and leave it thus for twenty
minutes. In the more expensive patterns of autoclave this regulation of
the safety valve is carried out automatically, the manometer being
fitted with an adjustable pointer which can be set to any required
pressure-temperature and so arranged that when the index of the
manometer coincides with the adjustable hand the safety valve is opened.
7. Extinguish the gas and allow the manometer index to fall to zero.
[Illustration: FIG. 30.--Chamberland's Autoclave.]