The cold junction of a thermo-couple may be connected by means
of copper wires to the voltmeter, although in some installations
of base metal couples, the wires forming the couple are themselves
extended to the voltmeter, making copper connections unnecessary.
From the foregoing, it may be seen that accurately to measure the
temperature of the hot end of a thermo-couple, we _must know the
temperature of the cold end_, as it is the _difference_ in the
temperatures that determines the voltmeter readings. This is absolutely
essential for precision, and its importance cannot be over-emphasized.
When pyrometers are used in daily operation, they should be checked
or calibrated two or three times a month, or even every week. Where
there are many in use, it is good practice to have a master pyrometer
of a rare metal couple, which is used only for checking up the
others. The master pyrometer, after calibrating against the melting
points of various substances, will have a calibration chart which
should be used in the checking operation.
It is customary now to send a rare metal couple to the Bureau of
Standards at Washington, where it is very carefully calibrated
for a nominal charge, and returned with the voltmeter readings
of a series of temperatures covering practically the whole range
of the couple. This couple is then used only for checking those
in daily use.
Pyrometer couples are more or less expensive, and should be cared
far when in use. The wires of the couple should be insulated from
each other by fireclay leads or tubes, and it is well to encase them
in a fireclay, porcelain, or quartz tube to keep out the furnace
gases, which in time destroy the hot junction. This tube of fireclay,
or porcelain, etc., should be protected against breakage by an
iron or nichrome tube, plugged or welded at the hot end. These
simple precautions will prolong the life of a couple and maintain
its precision longer.
Sometimes erroneous temperatures are recorded because the "cold
end" of the couple is too near the furnace and gets hot. This always
causes a temperature reading lower than the actual, and should be
guarded against. It is well to keep the cold end cool with water,
a wet cloth, or by placing it where coal air will circulate around
it. Best of all, is to have the cold junction in a box, together
with a thermometer, so that its temperature may definitely be known.
If this temperature should rise 20 deg.F. on a hot day, a correction of
20 deg.F. should be added to the pyrometer reading, and so on. In the
most up-to-date installations, this cold junction compensation is
taken care of automatically, a fact which indicates its importance.