As the intensity of the light emitted at the higher temperatures
becomes dazzling, it is found desirable to introduce a piece of red
glass in the eye piece at _R_. This also eliminates any question
of matching colors, or of the observer's ability to distinguish
colors. It is further of value in dealing with bodies which do
not radiate light of the same composition as that emitted by a
black body, since nevertheless the intensity of radiation of any
one color from such bodies increases progressively in a definite
manner as the temperature rises. The intensity of this one color
can therefore be used as a measure of temperature for the body
in question. Figures 124 to 126 show the way it is read.
CORRECTION FOR COLD-JUNCTION ERRORS
The voltage generated by a thermo-couple of an electric pyrometer is
dependent on the difference in temperature between its hot junction,
inside the furnace, and the cold junction, or opposite end of the
thermo-couple to which the copper wires are connected. If the
temperature or this cold junction rises and falls, the indications
of the instrument will vary, although the hot junction in the furnace
may be at a constant temperature.
A cold-junction temperature of 75 deg.F., or 25 deg.C., is usually adopted
in commercial pyrometers, and the pointer on the pyrometer should
stand at this point on the scale when the hot junction is not heated.
If the cold-junction temperature rises about 75 deg.F., where base metal
thermo-couples are used, the pyrometer will read approximately 1 deg.
low for every 1 deg. rise in temperature above 75 deg.F. For example, if the
instrument is adjusted for a cold-junction temperature of 75 deg., and
the actual cold-junction temperature is 90 deg.F., the pyrometer will
read 15 deg. low. If, however, the cold-junction temperature falls below
75 deg.F., the pyrometer will read high instead of low, approximately
1 deg. for every 1 deg. drop in temperature below 75 deg.F.
With platinum thermo-couples, the error is approximately 1/2 deg. for
1 deg. change in temperature.
CORRECTION BY ZERO ADJUSTMENT.--Many pyrometers are supplied with
a zero adjuster, by means of which the pointer can be set to any
actual cold-junction temperature. If the cold junction of the
thermo-couple is in a temperature of 100 deg.F., the pointer can be
set to this point on the scale, and the readings of the instrument
will be correct.