The eye soon becomes accustomed to estimating the tenths of the
divisions. If the paper is held as directed, two divisions below the
meniscus, one whole division is visible to correct the judgment. It is
not well to attempt to bring the meniscus exactly to a division mark
on the burette. Such readings are usually less accurate than those in
which the tenths of a division are estimated.]
CALIBRATION OF GLASS MEASURING DEVICES
If accuracy of results is to be attained, the correctness of all
measuring instruments must be tested. None of the apparatus offered
for sale can be implicitly relied upon except those more expensive
instruments which are accompanied by a certificate from the !National
Bureau of Standards! at Washington, or other equally authentic source.
The bore of burettes is subject to accidental variations, and since
the graduations are applied by machine without regard to such
variations of bore, local errors result.
The process of testing these instruments is called !calibration!.
It is usually accomplished by comparing the actual weight of water
contained in the instrument with its apparent volume.
There is, unfortunately, no uniform standard of volume which has been
adopted for general use in all laboratories. It has been variously
proposed to consider the volume of 1000 grams of water at 4°, 15.5°,
16°, 17.5°, and even 20°C., as a liter for practical purposes, and to
consider the cubic centimeter to be one one-thousandth of that volume.
The true liter is the volume of 1000 grams of water at 4°C.; but
this is obviously a lower temperature than that commonly found in
laboratories, and involves the constant use of corrections if taken as
a laboratory standard. Many laboratories use 15.5°C. (60° F.) as the
working standard. It is plain that any temperature which is deemed
most convenient might be chosen for a particular laboratory, but it
cannot be too strongly emphasized that all measuring instruments,
including burettes, pipettes, and flasks, should be calibrated at that
temperature in order that the contents of each burette, pipette,
etc., shall be comparable with that of every other instrument, thus
permitting general interchange and substitution. For example, it is
obvious that if it is desired to remove exactly 50 cc. from a solution
which has been diluted to 500 cc. in a graduated flask, the 50 cc.
flask or pipette used to remove the fractional portion must give
a correct reading at the same temperature as the 500 cc. flask.
Similarly, a burette used for the titration of the 50 cc. of solution
removed should be calibrated under the same conditions as the
measuring flasks or pipettes employed with it.